St. Jude Medical announces Class 1 recall of Optisure dual coil defibrillation leads

St. Jude Medical announced on Jan. 22 that the FDA had classified a previous voluntary recall of the company's Optisure dual coil defibrillation leads as a Class 1 device recall.

On Nov. 3, 2015, St. Jude Medical sent an advisory letter to doctors alerting them of issues with 447 Optisure leads, including 278 that were distributed in the U.S.

St. Jude Medical found that “during the manufacturing process of a limited number of Optisure leads, a trim technique to remove excess medical adhesive around the SVC shock coil may have introduced damage to the lead’s insulation,” according to an FDA notice.

There were no reports of lead malfunction or patient injury related to the issue, according to a St. Jude Medical news release.

The Optisure leads are used in conjunction with implantable cardioverter defibrillators. The leads were distributed in the U.S., Belgium, Switzerland, Germany, France, United Kingdom, India, Italy, Japan, Luxembourg, The Netherlands, Saudi Arabia and Sweden.

“A St. Jude Medical internal investigation found the probability that a lead was damaged as a result of the manufacturing variation to an extent that it could result in the inability to deliver appropriate high voltage therapy is very low and that any associated risks can be prevented with device reprogramming,” the company said in a news release. “The patients’ leads can also be monitored from home using the Merlin.net remote care system. The company has not received any reports of compromised performance of the impacted Optisure leads. St. Jude Medical is in the process of providing an updated advisory notice to physicians to further ensure physicians are aware of recommendations for managing their patients who may have been implanted with the impacted leads.”

Tim Casey,

Executive Editor

Tim Casey joined TriMed Media Group in 2015 as Executive Editor. For the previous four years, he worked as an editor and writer for HMP Communications, primarily focused on covering managed care issues and reporting from medical and health care conferences. He was also a staff reporter at the Sacramento Bee for more than four years covering professional, college and high school sports. He earned his undergraduate degree in psychology from the University of Notre Dame and his MBA degree from Georgetown University.

Around the web

Ron Blankstein, MD, professor of radiology, Harvard Medical School, explains the use of artificial intelligence to detect heart disease in non-cardiac CT exams.

Eleven medical societies have signed on to a consensus statement aimed at standardizing imaging for suspected cardiovascular infections.

Kate Hanneman, MD, explains why many vendors and hospitals want to lower radiology's impact on the environment. "Taking steps to reduce the carbon footprint in healthcare isn’t just an opportunity," she said. "It’s also a responsibility."