FDA confirms new Class I recall for heart valve delivery system due to safety risk

The U.S. Food and Drug Administration (FDA) has confirmed that a safety issue with certain accessories used to implant the Sapien M3 transcatheter mitral valve replacement (TMVR) system has been ruled a Class I recall. 

This is the agency’s most severe recall classification. It means using the devices in question could cause “serious adverse health consequences or death.”

This Class I recall only covers four specific lots of these devices: 67718561, 67804014, 67887043 and 67887044. Delivery systems distributed as part of those four lots should not be used to treat any patients going forward. However, all other delivery systems sold with the Sapien M3 TMVR system can be used as normal. 

The FDA first released information about this recall in October while the issue was still under investigation. As the agency explained at the time, four lots of the Commander M delivery systems sold with the Sapien M3 TMVR systems were being recalled due to an increased risk of the balloons bursting during inflation. Four serious injuries and one death were linked to the issue. 

Subscribe to Cardiovascular Business News

Key details about the Sapien M3 TMVR system

The Sapien M3 TMVR system first secured FDA approval in January for the treatment of mitral regurgitation (MR). It was designed for patients with symptomatic moderate-to-severe or severe MR who may not be good candidates for mitral valve surgery or transcatheter edge-to-edge repair. The device is also indicated to treat high-risk patients who present with symptomatic mitral valve dysfunction linked to mitral annular calcification. 

At TCT 2026 in San Diego, late-breaking data confirmed that the Sapien M3 TMVR system is associated with positive two-year outcomes for patients requiring TMVR. The device was confirmed to be both safe and effective in patients two years after treatment. Those findings were simultaneously published in JACC.[1]

Again, this recall only includes certain Commander M delivery systems. No issues have been identified with the actual Sapien M3 TMVR devices.

Michael Walter
Michael Walter, Managing Editor

Michael has more than 19 years of experience as a professional writer and editor. He has written at length about cardiology, radiology, artificial intelligence and other key healthcare topics.

Innovation at work

Nearly half of all American adults over the age of 20 have hypertension, and they face an increased risk of cardiovascular disease, renal events and cerebrovascular events. The recent rise of renal denervation has given cardiologists a new way to fight back against this trend.

Sponsored by Recor Medical

Curtiss Stinis, MD, has been performing TAVR for nearly two decades. He says the new policies from CMS that expand coverage for asymptomatic patients represent a "momentous step forward" for heart patients with severe aortic stenosis.

Subscribe to Cardiovascular Business News

Subscribe to Cardiovascular Business News